TUNKIA TK3300  miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench

Customization: Available
After-sales Service: Life Time
Warranty: 12 Months Excellent After Service

Products Details

  • Overview
  • Company Profile
Overview

Basic Info.

Model NO.
TK3300
Power
Electricity
Certification
SGS
Voltage Output
/
LCD Display
With LCD Display
Customized
Non-Customized
Accuracy
Class 0.01 or 0.02
3PH Voltage Measurement (direct)
288 V
3PH Voltage Measurement (indirect)
144 V
3PH Current Measurement (direct)
120 A
3PH Current Measurement (indirect)
12 A
Phase uncertainty
0.003° (class 0.01)
communication interface
RS485 and RS232
Power Supply
Built-in lithium battery power supply
Transport Package
Package for Exporting
Trademark
TUNKIA
Origin
China
Production Capacity
1000

Packaging & Delivery

Package Size
50.00cm * 50.00cm * 50.00cm
Package Gross Weight
10.000kg

Product Description

TK3300 Installed Three-Phase
Standard Energy Meter
TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench

1. Summary

The TK3300 miniature installed three-phase standard energy meter is a standard energy meter with the same physical dimensions as conventional installed three-phase energy meters. TK3300 is specially used for calibration or intermediate check of three-phase energy meter test bench or assembly line devices, and can complete testing items such as basic errors and multi-channel output consistency.
 

2. Features

  1. Accuracy class 0.01 and class 0.02 are optional.
  2. There are two types: direct connection and connection via transformer (indirect connection).
  3. Voltage measurement range: 288 V (direct connection), 144 V (indirect connection).
  4. Current measurement range: 120 A (direct connection), 12 A (indirect connection).
  5. Voltage and current support fully automatic switching of range.
  6. Phase uncertainty typical value: 0.003° (class 0.01).
  7. Lithium battery power supply to avoid the impact of PT power taking on measurement.
  8. The battery can last for 8 hours and the charging time is less than 2 hours.
  9. Optional manual charging or automatic charging. The automatic charging standard version requires an optional charging system.
  10. Supports RS485 and RS232 communication interface.
  11. An external display screen (optional) can be connected to display the measured power.

3. Applications

TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench
  Calibration of Three-Phase Energy Meter Device
  1. Install standard meters in batches to conduct calibration or intermediate check on the platform or assembly line equipment.
  2. Test items include basic errors, multi-channel output consistency, etc.


TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench

 Fully Automatic Intermediate Check
  1. Equipped with installed standard meters and automatic charging systems in batches
  2. Fully automatically complete the intermediate check of the assembly line equipment without manual participation.

4. Instrument Appearance


 Instrument Front Panel
picture S/N Function
TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench 1 Indicator lights, including pulse indicator lights, alarm indicators, and infrared indicators.
2 232 Interface, used to configure standard energy mete energy constants, etc.;
LCD Interface, used to connect the display screen;
3 Power button.
4 Power charging port.
5 Terminals.

5. Characteristics

Complies with Installed Meter Standard Dimensions
TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench
  1. The size, wiring terminals and pulse lamp position of the installed energy meter are consistent.
  2. It can be directly crimped to the meter position of the device without separate wiring, meeting the needs of automated detection.
 
Communication Interface
TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench
  1. Supports RS485 and RS232 communication.
  2. Set the energy pulse constant and read the power information.
  3. An external LCD display (optional) can be connected to display the measured power value in real time.
 
Automatic Charging System (optional)
TUNKIA TK3300 miniature Installed Three-Phase AC Reference Standard Energy Meter for Calibrating Energy Meters Test Bench
  1. The compact size does not affect the automatic operation of the assembly line.
  2. It is powered by lithium batteries when working.
  3. It can be automatically transferred to the cache area for charging when not working.
  4. It communicates and controls with the master control system through Ethernet.
  5. It has charging over-voltage, over-current and other fault alarm functions.
  6. It has automatic and manual start and stop functions.

6 Specifications

6.1 TK3300-X-D(Direct Connection)

6.1.1 Three-Phase Voltage Measurement

Range Measurement Uncertainty ( k = 2 )
( ppm*RD + ppm*RG ) [1]
Temperature Coefficient
@ (15~30)°C
(ppm*RD+ppm*RG) /°C
Class 0.02 Class 0.01 Class 0.02 Class 0.01
60 V 60 + 40 30 + 20 < 6 < 3
120 V 60 + 40 30 + 20 < 6 < 3
240 V 60 + 40 30 + 20 < 6 < 3
Note [1]: RD is the reading value, RG is the range value, the same below.
  1. Measuring range: 50 V~288 V, 7-digit display, automatic range switching

6.1.2 Three-Phase Current Measurement

Range Measurement Uncertainty ( k = 2 )
( ppm*RD + ppm*RG ) [1]
Temperature Coefficient
@ (15~30)°C
(ppm*RD+ppm*RG) /°C
Class 0.02 Class 0.01 Class 0.02 Class 0.01
50 mA 120 + 80 60 + 40 < 10 < 6
100 mA 60 + 40 30 + 20 < 6 < 3
200 mA 60 + 40 30 + 20 < 6 < 3
500 mA 60 + 40 30 + 20 < 6 < 3
1 A 60 + 40 30 + 20 < 6 < 3
2 A 60 + 40 30 + 20 < 6 < 3
5 A 60 + 40 30 + 20 < 6 < 3
10 A 60 + 40 30 + 20 < 6 < 3
20 A 60 + 40 30 + 20 < 6 < 3
50 A 60 + 40 30 + 20 < 6 < 3
100 A 60 + 40 30 + 20 < 6 < 3
  1. Measuring range: 5 mA~120 A, 7-digit display, automatic range switching
6.1.3 Frequency/Phase Measurement
Frequency Measuring Range 45 Hz~65 Hz
Measurement Uncertainty (k=2) 0.005%*RD
Phase Measuring Range 0~360°(I ≥ 50mA)
Measurement Uncertainty
(k=2)
Class 0.02 0.006°
Class 0.01 0.003°
 

6.1.4 AC Power/Energy Measurement

Current Range Power Factor Measurement Uncertainty(k=2)
Class 0.02 Class 0.01
100 mA ≤ I ≤120 A 0.5L~1~0.5C 0.02%*RD 0.01%*RD
50 mA ≤ I < 100 mA 1 0.02%*RD 0.01%*RD
0.5L~1~0.5C 0.04%*RD 0.02%*RD
5 mA ≤ I < 50 mA 1 0.04%*RD 0.02%*RD
0.5L~1~0.5C 0.08%*RD 0.04%*RD
  1. Power/energy measurement range: combination of AC voltage range and AC current range
  2. Power factor measurement range: -1.000 000…0.000 000…1.000 000
  3. Standard energy pulse output: high frequency full scale value corresponds to 60 kHz, low frequency full scale value corresponds to 6 Hz


 

6.2 TK3300-X-I(Indirect Connection)

6.2.1  Three-Phase Voltage Measurement

Range Measurement Uncertainty ( k = 2 )
( ppm*RD + ppm*RG ) [1]
Temperature Coefficient
@ (15~30)°C
(ppm*RD+ppm*RG) /°C
Class 0.02 Class 0.01 Class 0.02 Class 0.01
60 V 60 + 40 30 + 20 < 6 < 3
120 V 60 + 40 30 + 20 < 6 < 3
Note [1]: RD is the reading value, RG is the range value, the same below.
  1. Measuring range: 6 V~144 V, 7-digit display, automatic range switching
 

6.2.2 Three-Phase Current Measurement

Range Measurement Uncertainty ( k = 2 )
( ppm*RD + ppm*RG ) [1]
Temperature Coefficient
@ (15~30)°C
(ppm*RD+ppm*RG) /°C
Class 0.02 Class 0.01 Class 0.02 Class 0.01
10 mA 60 + 40 60 + 40 < 6 < 15
20 mA 80 + 60 40 + 30 < 3 < 6
50 mA 80 + 60 40 + 30 < 15 < 6
100 mA 60 + 40 30 + 20 < 6 < 3
200 mA 60 + 40 30 + 20 < 6 < 3
500 mA 60 + 40 30 + 20 < 6 < 3
1 A 60 + 40 30 + 20 < 6 < 3
2 A 60 + 40 30 + 20 < 6 < 3
5 A 60 + 40 30 + 20 < 6 < 3
10 A 60 + 40 30 + 20 < 6 < 3
  1. Measuring range: 1 mA~12 A, 7-digit display, automatic range switching
 

6.2.3 Frequency/Phase Measurement

Frequency Measuring Range 45 Hz~65 Hz
Measurement Uncertainty (k=2) 0.005%*RD
Phase Measuring Range 0~360°(I ≥ 50mA)
Measurement Uncertainty
(k=2)
Class 0.02 0.006°
Class 0.01 0.003°
 

6.2.4 AC Power/Energy Measurement

Current Range Power Factor Measurement Uncertainty(k=2)
Class 0.02 Class 0.01
100 mA ≤ I ≤ 12 A 0.5L~1~0.5C 0.02%*RD 0.01%*RD
10 mA ≤ I < 100 mA 1 0.02%*RD 0.01%*RD
0.5L~1~0.5C 0.04%*RD 0.02%*RD
1 mA ≤ I < 10 mA 1 0.04%*RD 0.02%*RD
0.5L~1~0.5C 0.08%*RD 0.04%*RD
  1. Power/energy measurement range: combination of AC voltage range and AC current range
  2. Power factor measurement range: -1.000 000…0.000 000…1.000 000
  3. Standard energy pulse output: high frequency full scale value corresponds to 60 kHz, low frequency full scale value corresponds to 6 Hz

7. General Specifications

Power Supply Built-in lithium battery power supply
Warm up time 30 minutes
Temperature Performance Operating temperature: 5°C~45°C;
Storage temperature: -10°C~55°C
Humidity Performance Working humidity: < 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C
Storage humidity: (20%~80%) R·H,no condensation.
Altitude < 3000 m
Weight About 1kg
Communication Interface RS232,RS485
Power Supply Built-in lithium battery power supply


 