- Overview
- Company Profile
Basic Info.
Model NO.
TK3100
Power
Electricity
Certification
SGS
Voltage Output
/
LCD Display
With LCD Display
Customized
Non-Customized
Accuracy
Class 0.02 or 0.05
3PH Voltage Measurement
50 V ~ 288 V
3PH Current Measurement
5 mA~120 A
Phase uncertainty
0.006°
communication interface
RS485 and RS232
Power Supply
Built-in lithium battery power supply
Transport Package
Package for Exporting
Trademark
TUNKIA
Origin
China
Production Capacity
1000
Packaging & Delivery
Package Size
20.00cm * 50.00cm * 30.00cm
Package Gross Weight
5.000kg
Product Description
TK3100 Installed Single-Phase
Standard Energy Meter
Standard Energy Meter
1. Summary
The TK3100 miniature installed single-phase standard energy meter is a standard energy meter with the same physical dimensions as conventional installed single-phase energy meters. It is specially used to verify the AC energy meter calibration devices or energy measurement calibration units in AC energy meter automated test system. It can complete testing items such as basic error and multi-channel output consistency.
2. Features
- Accuracy class 0.02 and class 0.05 are optional.
- Voltage measurement range: 50 V ~ 288 V.
- Current measurement range: 5 mA~120 A.
- Voltage and current support fully automatic switching of range.
- Phase measurement uncertainty typical value:0.006°.
- Lithium battery power supply to avoid the impact of PT power taking on measurement.
- The battery can last for 8 hours and the charging time is less than 2 hours.
- Optional manual charging or automatic charging. The automatic charging standard version requires an optional charging system.
- Supports RS485 communication and RS232 communication.
- An external display screen (optional) can be connected to display the measured power.
3. Applications
☆ Calibration of Three-Phase Energy Meter Device |
|
☆ Fully Automatic Intermediate Check |
|
4. Instrument Appearance
☆ Instrument Front Panel | ||
picture | S/N | Function |
1 | Indicator lights, including pulse indicator lights, alarm indicators, and infrared indicators. | |
2 | 232 Interface, used to configure standard energy mete energy constants, etc.; LCD Interface, used to connect the display screen; | |
3 | Power button. | |
4 | Power charging port. | |
5 | Terminals. |
5. Characteristics
☆ Complies with Installed Meter Standard Dimensions | |
|
☆ Communication Interface | |
|
☆ Automatic Charging System (optional) | |
|
6 Specifications
6.1 Voltage Measurement
Range | Measurement Uncertainty ( k = 2 ) ( ppm*RD + ppm*RG ) [1] | Temperature Coefficient @ (15~30)°C (ppm*RD+ppm*RG) /°C | ||
Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |
240 V | 120 + 80 | 60 + 40 | < 15 | < 6 |
Note [1]: RD is the reading value, RG is the range value, the same below. |
- Measuring range: 50 V~288 V, 7-digit display, automatic range switching
6.2 Current Measurement
Range | Measurement Uncertainty ( k = 2 ) ( ppm*RD + ppm*RG ) [1] | Temperature Coefficient @ (15~30)°C (ppm*RD+ppm*RG) /°C | ||
Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |
50 mA | 250 + 150 | 120 + 80 | < 20 | < 10 |
100 mA | 200 + 100 | 60 + 40 | < 15 | < 6 |
200 mA | 200 + 100 | 60 + 40 | < 15 | < 6 |
500 mA | 200 + 100 | 60 + 40 | < 15 | < 6 |
1 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
2 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
5 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
10 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
20 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
50 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
100 A | 200 + 100 | 60 + 40 | < 15 | < 6 |
- Measuring range: 5 mA~120 A, 7-digit display, automatic range switching
6.3 Frequency/Phase Measurement
Frequency | Measuring Range | 45 Hz~65 Hz | |
Measurement Uncertainty (k=2) | 0.005%*RD | ||
Phase | Measuring Range | 0~360°(I ≥ 50mA) | |
Measurement Uncertainty (k=2) | Class 0.05 | 0.012° | |
Class 0.02 | 0.006° |
6.4 AC Power/Energy Measurement
Current Range | Power Factor | Measurement Uncertainty(k=2) | |
Class 0.05 | Class 0.02 | ||
100 mA ≤ I ≤120 A | 0.5L~1~0.5C | 0.05%*RD | 0.02%*RD |
50 mA ≤ I < 100 mA | 1 | 0.05%*RD | 0.02%*RD |
0.5L~1~0.5C | 0.08%*RD | 0.04%*RD | |
5 mA ≤ I < 50 mA | 1 | 0.08%*RD | 0.04%*RD |
0.5L~1~0.5C | 0.15%*RD | 0.08%*RD |
- Power/energy measurement range: combination of AC voltage range and AC current range
- Power factor measurement range: -1.000 000…0.000 000…1.000 000
- Standard energy pulse output: high frequency full scale value corresponds to 60 kHz, low frequency full scale value corresponds to 6 Hz
7. General Specifications
Power Supply | Built-in lithium battery power supply |
Warm up time | 30 minutes |
Temperature Performance | Operating temperature: 5°C~45°C; Storage temperature: -10°C~55°C |
Humidity Performance | Working humidity: < 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity: (20%~80%) R·H,no condensation. |
Altitude | < 3000 m |
Weight | About 0.5 kg |
Communication Interface | RS232,RS485 |