TI1200 Precision DC Resistance Calibrator

Customization: Available
After-sales Service: Life Time
Warranty: 12 months Excellent After-sales Service

Products Details

  • Overview
  • Company Profile
Overview

Basic Info.

Model NO.
TI1200
Customized
Non-Customized
Accuracy
50 ppm
Interface
RS232; LAN; USB
Transport Package
Package for Exporting
Specification
5kg; 215 mm (W) x 252 mm (D) x 88 mm(H),
Trademark
TUNKIA
Origin
China

Packaging & Delivery

Package Size
28.00cm * 30.00cm * 15.00cm
Package Gross Weight
10.000kg

Product Description

TI1200 Precision DC Resistance Calibrator
TI1200 Precision DC Resistance Calibrator

1. Summary

The TI1200 is a high precision DC resistance source with wide range, it is applied for calibrating ohmmeter on-site, e.g, industrial production lines, etc.

2. Features

  1. Resistance output: 0 Ω ~ 1100 MΩ
  2. Accuracy: 50 ppm
  3. Shortcut output keys
  4. Programmable calibration schemes
  5. RS232, LAN and USB communication interfaces

3. Application

  1. Calibrate DC resistance meters
  2. Calibrate digital multimeters
  3. Calibrate DC bridges, microhmmeters.

4. Features


Convenient operation
TI1200 Precision DC Resistance Calibrator
  1. The front panel has 10 shortcut keys that can output the corresponding value on the screen with one key.
 
Calibration scheme setting
TI1200 Precision DC Resistance Calibrator
  1. Multiple calibration schemes can be edited and stored for the calibration of multimeter with different specifications.

5. Specifications

Range Resistance Output Resolution Accuracy
±(ppm*output+Ω)
Allowable Current
10 Ω 0 Ω ~ 11.00000 Ω 10 μΩ 50 + 0.01 1 mA ~ 150 mA
30 Ω 10.00000 Ω ~ 33.00000 Ω 10 μΩ 40 + 0.015 1 mA ~ 150 mA
100 Ω 30.0000 Ω ~ 110.0000 Ω 100 μΩ 35 + 0.015 1 mA ~ 80 mA
300 Ω 100.0000 Ω ~ 330.0000 Ω 100 μΩ 35 + 0.02 1 mA ~ 40 mA
1kΩ 0.300000 kΩ ~ 1.100000 kΩ 1 mΩ 35 + 0.02 1 mA ~ 20 mA
3 kΩ 1.000000 kΩ ~ 3.300000 kΩ 1 mΩ 35 + 0.15 0.1 mA ~ 6 mA
10 kΩ 3.00000 kΩ ~ 11.00000 kΩ 10 mΩ 35 + 0.15 0.1 mA ~ 2 mA
30 kΩ 10.00000 kΩ ~ 33.00000 kΩ 10 mΩ 35 + 2 10 μA ~ 0.6 mA
100 kΩ 30.0000 kΩ ~ 110.0000 kΩ 100 mΩ 35 + 3 10 μA ~ 0.2 mA
300 kΩ 100.0000 kΩ ~ 330.0000 kΩ 100 mΩ 40 + 30 1 μA ~ 60 μA
1 MΩ 0.300000 MΩ ~ 1.100000 MΩ 1 Ω 40 + 30 1 μA ~ 20 μA
3 MΩ 1.000000 MΩ ~ 3.300000 MΩ 1 Ω 50 + 300 0.25 μA ~ 6 μA
10 MΩ 3.00000 MΩ ~ 11.00000 MΩ 10 Ω 200 + 500 0.25 μA ~ 2 μA
30 MΩ 10.00000 MΩ ~ 33.00000 MΩ 10 Ω 400 + 5000 25 nA ~ 600 nA
100 MΩ 30.0000 MΩ ~ 110.0000 MΩ 100 Ω 1000 + 6000 25 nA ~ 200 nA
300 MΩ 100.0000 MΩ ~ 330.0000 MΩ 100 Ω 5000 + 200000 2.5 nA ~ 60 nA
1000 MΩ 300.000 MΩ ~ 1100.000 MΩ 1 kΩ 20000 + 600000 1 nA ~ 20 nA
  1. Output range: 0 Ω ~ 1100 MΩ, 7 digits display.
  2. Adjustment fineness: 0.002%*RG.
  3. Manual or automatic range switch.
  4. Four-terminals or Two-terminals.
  5. Reverse connect protection, Overload protection

6. General Specifications

Power supply AC (220±22) V, (50±2) Hz
Maximum power consumption 100 VA
Warm up time Twice the time since last warmed up, to a maximum of 30 minutes.
Temperature performance 0°C ~ 55°C, 30%R·H ~ 80%R·H, non-condensing
Humidity performance -30°C ~ 70°C, 10%R·H ~ 90%R·H, non-condensing
Communication interface USB×1,LAN×1,RS232×1
Dimensions 215 mm (width) x 252 mm (depth) x 88 mm (height), excluding feets.
Display 3.5-inch color LCD
Weight About 5 kg

Company Profile
 

TI1200 Precision DC Resistance Calibrator

TI1200 Precision DC Resistance CalibratorTI1200 Precision DC Resistance Calibrator

 

Contact us

Please feel free to give your inquiry in the form below We will reply you in 24 hours