- Overview
- Product Parameters
- Packaging & Shipping
Basic Info.
Packaging & Delivery
To be negotiated ( > 100 Pieces)
Product Description
* Sources and measures DC voltage, current, resistance, RTDs and thermocouples
* Extend pressure measurement range with any Fluke 750P and 700P Series Pressure Modules
* Includes 24V DC transmitter loop power supply
* Measures 4 to 20mA loop current
* Includes automated switch-test function
* Accepts ITS-90 coefficients for accurate SPRT measurements
* Compatible with MET/CAL Calibration Software
* Includes certificate of calibration traceable to national standards (optional A2LA accredited calibration available upon
request)
* Sources and measures DC voltage, current, resistance, RTDs and thermocouples
* Extend pressure measurement range with any Fluke 750P and 700P Series Pressure Modules
* Includes 24V DC transmitter loop power supply
* Measures 4 to 20mA loop current
* Includes automated switch-test function
* Accepts ITS-90 coefficients for accurate SPRT measurements
* Compatible with MET/CAL Calibration Software
* Includes certificate of calibration traceable to national standards (optional A2LA accredited calibration available upon
request)
calibrate the temperature sensor itself, whether an RTD or thermocouple. Ignoring the sensor can be a mistake, because temperature sensors are responsible for more than 75% of the output errors in temperature transmitters. You can calibrate the temperature sensor individually, or calibrate both the sensor and the transmitter as a system using a dry-block calibrator such as the Fluke Calibration 914X Series Field Metrology Wells. The Field Metrology Wells were designed specifically with process calibration in mind-optimizing speed to temperature, portability, stability and accuracy. The 7526A and a 914X dry-block calibrator make a perfect combination for calibrating just about any temperature transmitter.
Standard Interface | RS-232, IEEE-488 (GPIB) |
| Operating: 0 to 50°C Calibration (tcal): 18 to 28°C Storage: -20 to 70°C |
Electromagnetic Compatibility | CE: Conforms to EN61326 (operation in controlled EM environments) |
Temperature Coefficient | For temperatures outside tcal 5°C is 10% of the 90-day specification (or 1 year if applicable) per °C |
| Operating: <80% to 30°C <70% to 40°C <40% to 50°C |
Altitude | Operating: 9,800ft (3,000m) max Non-operating: 40,000ft (12,200m) max |
Safety | EN/IEC 61010-1:2010 3rd Edition, UL 61010-1:2012, CAN/CSA 22.2 No. 61010-1-12 |
Analog Low Isolation | 20V |
Line Power | 120V~: 100 to 120V 240V~: 220 to 240V |
Line Frequency | 47 to 63Hz |
Line Voltage Variation | ±10% about setting |
Power Consumption | 15VA maximum |
| Height: 5.75in (14.6cm) Width: 17.5in (44.5cm) Depth: 11.75in (29.8cm) |
Weight (without options) | 9.35lbs (4.24kg) |
As an important international instrument sales platform, our company has the authorization qualifications of many international first-line brands. Relying on the advanced digital information management platform and international procurement platform, we provide more than 100,000 instruments, industrial MRO products, industrial products and spare parts to global customers. Our business scope: industrial testing, non-destructive testing, gas testing, electrical testing, laboratory testing, water quality testing, environmental testing, medical testing...
So far, we have become an important partner of many global top 500 companies and well-known local Chinese companies. Create an efficient supply chain system to provide customers with strong supply guarantee.