Etx-1825 Multifunctional Process Calibrator with 0.05 Basic Accuracy

Customization: Available
After-sales Service: Lifelong
Warranty: 1 Year

Products Details

  • Overview
  • Product Description
  • Packaging & Shipping
  • Company Profile
Overview

Basic Info.

Model NO.
ETX-1825
Power
Multifunction Temperature Calibrator
Thermocouple (TC)
K
Resistance-Temperature Detector (RTD)
Pt100(385)
LCD Display
With LCD Display
Customized
Customized
Power Supply
6 PCS of 7# Ni-MH Batteries or Power Adapter
Size and Weight
200mm*96mm*47mm, 650g
Calibration
Dcv Dci, Dci, Frequency, Tc (Refer to Manual)
Accessories
Test Leads, Battery, Manual, Power Adapter
Basic Accuracy
0.05%
Product Name
Etx-1825 Multifunctional Process Calibrator with 0
Transport Package
Carton
Specification
28x28x11cm
Trademark
CKT
Origin
China
HS Code
9031809090
Production Capacity
2000 Sets Per Month

Packaging & Delivery

Package Size
28.00cm * 28.00cm * 11.00cm
Package Gross Weight
3.000kg
Lead Time
10 days (1 - 1 Pieces)
To be negotiated ( > 1 Pieces)

Product Description

Product Description

https://ckt2017.en.made-in-china.com/product-list-1.html

Brief Introduction
A split-screen display. The upper display allows you to measure volts, current (with loop power), resistance and continuity test. The lower display allows you to measure and source volts, current, resistance temperature detectors (RTD), thermocouples (TC), frequency, and ohms.
Calibrates a transmitter using the split-screen.
The measure of thermocouple provides automatic reference junction
Automatic and manual cold end temperature compensation for TC.
Manual stepping and automatic stepping and ramping.
Super-strong protection:The waterproof grade is IP67. Automatic protection against the signal terminals connect to the 220V.
Support for PC communication

Summary of Source and Measure Functions

Function Measure Source
DC V 0~50V (the upper display ± 30V) 0~10V
DC mA 0~24mA (the upper display ± 24mA) 0~24mA
Frequency 1.000Hz ~ 100.00kHz 0.00Hz~20.000kHz
Resistance 0~3200Ω 0~3200Ω
DC mV 0~100mV (the upper display ± 200mV) 0~100mV
RTD Pt100, Pt1000, Cu50, Cu100 Pt100, Pt1000, Cu50, Cu100
TC E, J, K, T, B, R, S, N E, J, K, T, B, R, S, N
Others Loop supply, Step, Ramp, Dual display

Specifications

Specifications are based on a one year calibration cycle and apply from +18 C to +28 C unless stated otherwise. All specifications assume a 10 minute warmup period.

DC Voltage Measurement

Range Maximum measurement range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
28V
(upper display)
-33V~33V 0.001V 0.05+2 0.025+2
200mV
(upper display)
-80mV~80mV 0.001mV 0.05+20 0.025+20
-200mV~200mV 0.01mV 0.05+2 0.025+2
50V
(lower display)
-1V~60V 0.001V 0.05+2 0.02+2
100mV
(lower display)
-15mV~80mV 0.001mV 0.05+20 0.02+20
80mV~125mV 0.01mV 0.05+2 0.02+2
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC.
Input resistance: >1MΩ.

DC Voltage Source

Range Maximum output range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
100mV -15mV~99.999mV 0.001mV 0.05+20 0.02+20
100mV~125mV 0.01mV 0.05+2 0.02+2
10V 0~11V 0.001V 0.05+2 0.02+2
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC.
Maximum load: 1mA or 1kΩ (It should be based on the lower load.)
 

DC mA measurement

Range Maximum measurement range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
20mA
(upper display)
-24mA~24mA 0.001mA 0.05+2 0.025+2
20mA
(loop of upper display)
0~24mA 0.001mA 0.05+2 0.025+2
20mA
(lower display)
0~24mA 0.001mA 0.05+2 0.02+2
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC..

Input resistance:<100Ω.

DC mA source

Range Maximum output range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
20mA 0~24mA 0.001mA 0.05+2 0.02+2
20mA
(Transducer simulation)
0~24mA 0.001mA 0.05+2 0.02+2
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC.
Maximum load voltage: 20V, equivalent to voltage of 20mA on 1000Ω load resistance.

Resistance Measurement (upper display)

Range Maximum output range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
400Ω 0~440Ω 0.1Ω 0.05+2
3200Ω 420Ω~3300Ω
On-off test 0~200Ω 0.05+1
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC.
Maximum load voltage: 20V, equivalent to voltage of 20mA on 1000Ω load resistance.

Resistance measurement (lower display)

Range Maximum measurement range Resolution Accuracy(Ω)
ETX-1825 ETX-2025
Two-wire, three-wire Four-wire Two-wire, three-wire Four-wire
400Ω 0~440Ω 0.01Ω 0.25 0.15 0.15 0.10
3200Ω 420Ω~3600Ω 0.1Ω 1.5 1.0 1.0 0.5
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC.
Exciting current during measurement:
400Ω:0.4mA±10%;
3200Ω:0.2mA±10%.
Two-wire: Does not include lead resistance.
Three-wire: Assumes matched leads with a total resistance not exceeding 25Ω.

Resistance source

Range Maximum output range Resolution External exciting current Accuracy (Ω)
ETX-1825 ETX-2025
400Ω 0~440Ω 0.01Ω 0.40mA~3.30mA 0.25 0.15
3200Ω 400~3600Ω 0.1Ω 0.1mA~0.6mA 1.0 0.50
-10ºC~18ºC,+28ºC~55ºC temperature coefficient,±0.005%FS/ºC.

Frequency measurement

Range Maximum measurement range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
100Hz 1~99.999Hz 0.001Hz 0.02+1 0.01+1
1000Hz 100~999.99Hz 0.01Hz
10kHz 1k~9.9999kHz 0.1Hz
100kHz 10k~99.999kHz 1Hz
Sensitivity: 10Hz~10kHz,Vp-p≥1V; rest:Vp-p≥2V.
Wave form: Square wave. 5 counting points should be added to errors of other wave forms. Commercial frequency can be measured directly. 

Frequency source

Range Maximum output range Resolution Accuracy ( % of reading + counts)
ETX-1825 ETX-2025
200Hz 0~200Hz 0.01Hz 0.02+1 0.01+1
2000Hz 0~2000Hz 0.1Hz
20kHz 0~20kHz 1Hz
Output amplitude: ≥4.5Vp-p;
Wave form: Square wave

Temperature, thermocouples

Graduation Range Resolution Accuracy (ºC)
ETX-1825 ETX-2025
J -200ºC~0ºC
0ºC~1200ºC
0.1ºC 1.5ºC
1.0ºC
1.0ºC
0.7ºC
K -200ºC~0ºC
0ºC~1370ºC
0.1ºC 1.8ºC
1.2ºC
1.2ºC
0.8ºC
T -200ºC~0ºC
0ºC~400ºC
0.1ºC 1.8ºC
1.2ºC
1.2ºC
0.8ºC
E -200ºC~0ºC
0ºC~950ºC
0.1ºC 1.5ºC
1.0ºC
0.9ºC
0.7ºC
R -20ºC~0ºC
0ºC~500ºC
500ºC~1750ºC
0.1ºC 4ºC
2.5ºC
2ºC
2.5ºC
1.8ºC
1.4ºC
S -20ºC~0ºC
0ºC~500ºC
500ºC~1750ºC
0.1ºC 4ºC
2.5ºC
2ºC
2.5ºC
1.8ºC
1.5ºC
B 600ºC~800ºC
800ºC~1000ºC
1000ºC~1800ºC
0.1ºC 3.5ºC
2.5ºC
2ºC
2.2ºC
1.8ºC
1.4ºC
N -200ºC~0ºC
0ºC~1300ºC
0.1ºC 2.0ºC
1.2ºC
1.5ºC
0.9ºC
Errors of cold-junction compensation are not included in the table.
Accuracy of cold-junction compensation: 1.5ºC

Temperature, RTD

Graduation Range Resolution Accuracy (ºC)
ETX-1825 ETX-2025
2-wire
3-wire
4-wire Output 2-wire
3-wire
4-wire Output
Pt100 -200ºC~850ºC 0.1ºC 0.7 0.4 0.7 0.4 0.3 0.3
Pt1000 -200ºC~650ºC 0.4 0.3 0.3 0.3 0.15 0.15
Cu50 -50ºC~150ºC 1.2 0.8 0.8 0.8 0.5 0.5
Cu100 -50ºC~150ºC 0.7 0.4 0.4 0.4 0.25 0.25
As for exciting current during measurement, please refer to resistance measurement function.
As for allowable external exciting current during output, please refer to resistance output function.
2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance not exceeding 25Ω.
 

 
Packaging & Shipping

https://ckt2017.en.made-in-china.com/company-Changzhou-Chuangkai-Electronic-Co-Ltd-.html

Etx-1825 Multifunctional Process Calibrator with 0.05 Basic Accuracy
Company Profile

https://ckt2017.en.made-in-china.com/contact-info.html

Chuangkai Electronic Co., Ltd is leading company specialized in producing, developing & selling various types of measuring instruments and testing equipment, such as temperature data logger, LCR meter, DC resistance tester (micro ohm meter), digital power meter, power supply, transformer tester, battery tester, programmable DC electronic load, charger, temperature controller, tachometer, high voltage signal-amplifier, power amplifier, etc.

FAQ: 
Q. How is your company's product after-sales service system?
A. The product warranty period is 12 months, we can provide lifetime maintenance. We have professional pre-sales and after-sales teams that can respond to you within 24 hours to resolve technical issues. We provide one-to-one after-sales service.

Q: Can you customize it?
A: Yes, we accept customization according to customer's specific requirements.

Q. How long is your lead time? If I want to customize instruments, how long will it take?
A. When order quantity is 1-10 units, we have ready stock goods, we can ship out within 3-5 working days after receipt of deposit; products required new production of common materials, usually the lead time is about 10-15 days; products requiring new production of special and rare materials, usually the lead time is about 20-30 days.

Q. What payment method do you accept?
A. Our company accept payment method of T/T (bank transfer), Western Union, PayPal and so on.

Q. What package do you use for your products? What shipping method do you use?
A.1. Our company use standard exporting carton for packing. 2. We ship via express courier (door to door service), by air, and by sea according to customers' requirements.

For more questions, please feel free to send inquiries to us.

 

Contact us

Please feel free to give your inquiry in the form below We will reply you in 24 hours